opentrons-thermocycler

Opentrons Thermocycler Module - automated PCR thermal cycling with independent block (4-99°C) and lid (37-110°C) temperature control, profile execution, and auto-sealing lid support (GEN2) for high-throughput molecular biology workflows

jkitchin · 已有静态检查记录

安装与使用

运行命令前先阅读来源与依赖说明。复制命令不会自动安装或执行。

npx skills add jkitchin/knowhow --skill opentrons-thermocycler

技能说明


name: opentrons-thermocycler description: Opentrons Thermocycler Module - automated PCR thermal cycling with independent block (4-99°C) and lid (37-110°C) temperature control, profile execution, and auto-sealing lid support (GEN2) for high-throughput molecular biology workflows allowed-tools: ["*"]

Opentrons Thermocycler Module

Overview

The Opentrons Thermocycler Module automates PCR thermal cycling with precise, independent control of block temperature (4-99°C) and heated lid (37-110°C). Execute complex temperature profiles with automatic cycling, integrate with liquid handling for complete PCR setup automation, and use auto-sealing lids (GEN2) for walk-away workflows.

Core value: Fully automate PCR setup and cycling. Load samples, dispense reagents, seal plates (GEN2), run thermal profiles, and collect products—all without manual intervention.

When to Use

Use the Thermocycler skill when:

  • Running PCR amplification protocols
  • Performing qPCR or RT-PCR reactions
  • Automating DNA/RNA thermal cycling workflows
  • Executing multi-step temperature incubations
  • Integrating thermal cycling with automated liquid handling
  • Setting up high-throughput PCR screening

Don't use when:

  • Simple temperature control needed (use Temperature Module for 4-95°C range)
  • Shaking/mixing required during incubation (use Heater-Shaker Module)
  • Temperature outside 4-99°C range (block) or 37-110°C (lid)

Quick Reference

OperationMethodKey Parameters
Load moduleprotocol.load_module()"thermocyclerModuleV2" or "thermocyclerModuleV1"
Open lidopen_lid()-
Close lidclose_lid()-
Set lid temperatureset_lid_temperature()celsius (37-110)
Deactivate liddeactivate_lid()-
Set block temperatureset_block_temperature()celsius (4-99), hold_time, block_max_volume
Execute profileexecute_profile()steps, repetitions, block_max_volume
Deactivate blockdeactivate_block()-

Platform Compatibility

Both Opentrons Flex and OT-2

Module Generations

  • GEN1 - Original thermocycler, OT-2 only
  • GEN2 - Improved plate lift mechanism, auto-sealing lid support

API compatibility: Both generations support identical API methods

Deck Position

OT-2: Spans multiple deck slots (typically slots 7, 8, 10, 11)

Flex: Dedicated thermocycler position

Loading:

# No deck slot specified - thermocycler has fixed position
tc_mod = protocol.load_module("thermocyclerModuleV2")

Loading the Module

from opentrons import protocol_api

metadata = {'apiLevel': '2.19'}

def run(protocol: protocol_api.ProtocolContext):
    # Load Thermocycler Module (no location - fixed position)
    tc_mod = protocol.load_module("thermocyclerModuleV2")

    # Load PCR plate
    pcr_plate = tc_mod.load_labware("opentrons_96_wellplate_200ul_pcr_full_skirt")

Module versions:

  • "thermocyclerModuleV1" - GEN1
  • "thermocyclerModuleV2" - GEN2 (recommended)

Lid Control

Basic Lid Operations

# Open lid for pipette access
tc_mod.open_lid()

# Close lid for thermal cycling
tc_mod.close_lid()

Important for gripper (Flex):

  • Always open lid before gripper operations
  • Close lid only after plate is seated on thermocycler

Heated Lid

The heated lid prevents condensation during thermal cycling:

# Set lid temperature (blocks until reached)
tc_mod.set_lid_temperature(celsius=105)

# Run thermal profile
# ... cycling ...

# Turn off lid heater
tc_mod.deactivate_lid()

Lid temperature range: 37-110°C

Typical settings:

  • Standard PCR: 105°C
  • RT-PCR: 100-105°C
  • Custom protocols: Match or exceed highest block temperature + 5-10°C

Note: Lid temperature is independent of block temperature and profiles.

Block Temperature Control

Basic Block Temperature

# Set block temperature and wait
tc_mod.set_block_temperature(
    temperature=95,
    hold_time_seconds=180,  # Hold for 3 minutes
    block_max_volume=50     # Volume in wells (µL)
)

# Block automatically maintains temperature for hold_time

Block temperature range: 4-99°C

Parameters:

  • temperature - Target temperature in °C (required)
  • hold_time_seconds - Duration to hold at temperature
  • hold_time_minutes - Alternative duration in minutes
  • block_max_volume - Sample volume in µL (default: 25µL)

Hold Time Behavior

With hold_time: Protocol waits at temperature for specified duration

# Wait 5 minutes at 95°C
tc_mod.set_block_temperature(
    temperature=95,
    hold_time_minutes=5,
    block_max_volume=50
)
# Protocol continues after 5 minutes

Without hold_time: Set temperature and continue immediately

# Set to 4°C and proceed (no hold)
tc_mod.set_block_temperature(temperature=4)

# Can perform other operations while maintaining 4°C

Block Max Volume

Specify sample volume for improved temperature accuracy:

# 50µL reactions
tc_mod.set_block_temperature(
    temperature=72,
    hold_time_minutes=10,
    block_max_volume=50  # Optimizes heating/cooling for 50µL
)

Default: 25µL if not specified

Why it matters: Algorithm adjusts for thermal mass of liquid to ensure accurate sample temperature.

Temperature Profiles

Automate repeated temperature cycles for PCR:

Defining a Profile

# PCR profile: Denature → Anneal → Extend
pcr_profile = [
    {"temperature": 95, "hold_time_seconds": 30},  # Denaturation
    {"temperature": 57, "hold_time_seconds": 30},  # Annealing
    {"temperature": 72, "hold_time_seconds": 60}   # Extension
]

Profile structure: List of dictionaries with temperature and hold_time_seconds

Executing a Profile

# Run profile for 30 cycles
tc_mod.execute_profile(
    steps=pcr_profile,
    repetitions=30,
    block_max_volume=50
)

**Parameters: