opentrons-magnetic-block
Opentrons Magnetic Block for Flex - unpowered magnetic bead separation using gripper-based labware movement with high-strength neodymium magnets for DNA/RNA purification, immunoprecipitation, and bead-based workflows
jkitchin · Static checks on file
Installation
Read the source and dependencies before running a command. Copying does not install or execute anything.
npx skills add jkitchin/knowhow --skill opentrons-magnetic-blockAbout this skill
name: opentrons-magnetic-block description: Opentrons Magnetic Block for Flex - unpowered magnetic bead separation using gripper-based labware movement with high-strength neodymium magnets for DNA/RNA purification, immunoprecipitation, and bead-based workflows allowed-tools: ["*"]
Opentrons Magnetic Block
Overview
The Opentrons Magnetic Block is an unpowered, 96-well magnetic separator exclusively for Opentrons Flex. It uses high-strength neodymium magnets to pull magnetic beads to the side of wells, enabling supernatant removal for DNA/RNA purification, protein cleanup, immunoprecipitation, and other bead-based separation protocols.
Core value: Automate magnetic bead-based purification workflows without manual plate transfers. Combine gripper-based plate movement with magnetic separation for fully automated DNA extraction, PCR cleanup, and bead-based assays.
Platform: Opentrons Flex only (not compatible with OT-2)
When to Use
Use the Magnetic Block skill when:
- Performing DNA/RNA extraction with magnetic beads
- PCR product cleanup (AMPure, SPRIselect beads)
- Magnetic bead-based library preparation (NGS)
- Protein purification or immunoprecipitation with magnetic beads
- Cell sorting or enrichment with magnetic particles
- Any protocol requiring magnetic bead separation
Don't use when:
- Working with OT-2 robot (Magnetic Block is Flex-only)
- Need vertical bead movement within solution (block only pulls to side)
- Require heated magnetic separation
- Working without Opentrons Flex gripper
Quick Reference
| Operation | Method | Key Parameters |
|---|---|---|
| Load module | protocol.load_module() | "magneticBlockV1", location |
| Move plate to block | protocol.move_labware() | labware, mag_block, use_gripper=True |
| Wait for separation | protocol.delay() | minutes (typically 2-5) |
| Pipette supernatant | pipette.transfer() | source, dest |
| Move plate off block | protocol.move_labware() | labware, new_location, use_gripper=True |
Platform Requirements
Opentrons Flex only
- API version: 2.15 or later (Magnetic Block introduced)
- Gripper required: All plate movement uses gripper
- Robot type: Must specify
"robotType": "Flex"
Key Characteristics
Unpowered Design
Important: The Magnetic Block is completely unpowered:
- No electronic control
- No on/off switching
- Always magnetic when plate is present
- Robot and Opentrons App are not aware of magnetic state
Implication: All control is manual through:
- Physical plate positioning (gripper movement)
- Timed delays for bead separation
- Pipetting operations
Loading the Module
from opentrons import protocol_api
metadata = {'apiLevel': '2.19'}
requirements = {"robotType": "Flex", "apiLevel": "2.19"}
def run(protocol: protocol_api.ProtocolContext):
# Load Magnetic Block
mag_block = protocol.load_module(
module_name="magneticBlockV1",
location="D1"
)
# Load compatible labware
mag_plate = mag_block.load_labware(
name="biorad_96_wellplate_200ul_pcr"
)
Compatible deck slots: Any Flex slot (A1-D3) Recommended: D1, D2, or D3 (bottom row for workflow efficiency)
Compatible Labware
The Magnetic Block works with 96-well plates that fit the magnetic footprint:
Common compatible labware:
biorad_96_wellplate_200ul_pcrnest_96_wellplate_100ul_pcr_full_skirtopentrons_96_wellplate_200ul_pcr_full_skirt- Other 96-well PCR plates with standard footprint
Important:
- Labware must be grippable (compatible with Flex gripper)
- Wells must align with magnet positions
- Standard 96-well footprint required
Basic Magnetic Separation Workflow
Step 1: Load Sample Plate
# Start with sample plate on deck (not on magnetic block yet)
sample_plate = protocol.load_labware("biorad_96_wellplate_200ul_pcr", "C1")
# Add magnetic beads and binding buffer
pipette.transfer(10, bead_stock, sample_plate.wells())
pipette.transfer(100, binding_buffer, sample_plate.wells(), mix_after=(5, 80))
Step 2: Incubate for Binding
# Incubate to allow beads to bind target (DNA/RNA/protein)
protocol.delay(minutes=5)
# Optional: Mix during incubation (use Heater-Shaker)
Step 3: Move to Magnetic Block
# Move plate to magnetic block using gripper
protocol.move_labware(
labware=sample_plate,
new_location=mag_block,
use_gripper=True
)
Step 4: Wait for Bead Separation
# Allow beads to collect at side of wells
protocol.delay(minutes=3)
# Beads are now held at well edges by magnets
Step 5: Remove Supernatant
# Pipette supernatant while beads are held by magnets
pipette.transfer(
volume=110,
source=sample_plate.wells(),
dest=waste.wells(),
new_tip="always"
)
# Beads remain in wells, held by magnetic field
Step 6: Move Off Magnetic Block
# Move plate off magnets for bead resuspension
protocol.move_labware(
labware=sample_plate,
new_location="C2",
use_gripper=True
)
# Beads are now free to resuspend
Step 7: Wash and Repeat
# Add wash buffer
pipette.transfer(150, wash_buffer, sample_plate.wells())
# Mix to resuspend beads
pipette.mix(repetitions=5, volume=100, location=sample_plate.wells())
# Return to magnetic block
protocol.move_labware(sample_plate, mag_block, use_gripper=True)
# Wait for separation
protocol.delay(minutes=2)
# Remove supernatant
pipette.transfer(150, sample_plate.wells(), waste.wells(), new_tip="always")
Step 8: Elute
# Move off magnets
protocol.move_labware(sample_plate, "C2", use_gripper=True)
# Add elution buffer
pipette.transfer(50, elution_buffer, sample_plate.wells(), mix_after=(5, 30))
# Incubate
protocol.delay(minutes=2)
# Final magnetic separation
protocol.move_labware(sample_plate, mag_block, use_gripper=True)
protocol.delay(minutes=2)
#